6 Technical data
Oxygen analyzer
PMA1000
PMA1000L
Part-No.
08 A 1000
08 A 1010
Sample gas
O
2
Measuring ranges
4 linear measuring ranges, 2 of those free selectable,
lowest span 1%
Factory default 0-1, 0-10, 0-30 and 0-100 vol% O
2
, zero
suppression applicable
Limit of Detection (LOD)**
0.02 vol%
Response time* for 90% FSD
< 3 seconds at 60 Nl/h air
Zero-point offset (drift)
< 0.06 vol% O
2
in 72 hours
Linearity error
< ±1 vol% O
2
Accuracy after calibration**
Deviation ±1 % of full scale or 0.02 vol% O
2
, whichever
value is greater.
Reproducibility**
< ±0.01 vol%
Flow rate of sample gas
25-60 Nl/h air
Influence of sample gas flow
Variation in gas flow between 25-60 Nl/h air will cause
a deviation of < 0.1 vol% O
2
Sample gas inlet pressure
0.6 - 1.6 bar
Sample gas outlet pressure
Recommendation: discharge freely into atmosphere
(requires higher pressure at the analyzer inlet com-
pared to the outlet)
Influence of sample gas pressure
< 1 % of full scale within the range of 0.6 to 1.6 bar abs.
with activated pressure compensation
Sample gas temperature and
characteristics
0 °C up to +50 °C (+32 °F up to +122 °F); dry, oil- and
dust free gas , avoid temperature dropping below dew
point
O
2
transducer temperature
Fixed at +55 °C (+131 °F)
Ambient temperature
0 °C up to +50 °C (+32 °F up to +122 °F), avoid tempera-
ture dropping below dew point
Influence of ambient
temperature
< 1 % of full scale
Display
7“ resistive touchscreen
Output signals
Adjustable 0-20 mA / 4-20 mA, max 500 Ohms burden,
short-circuit proof, electrically isolated, Modbus TCP/IP,
AK protocol TCP/IP
Relay outputs
4 x relay output (1 x status,
1 x Cal-mode, 1 x pump
control, 1 x Cal-error),
contacts: 24 V / 2 A,
1 x change-over contact,
potential-free
2 x relay output (1 x status,
1 x Cal-mode),
contacts: 24 V / 2 A,
1 x change-over contact,
potential-free
Binary outputs
8 x High-Side output 24 V,
max. 100 mA
(2 x operating parameters,
2 x measuring range
feedback, 4 x valve
control)
4 x High-Side output 24 V,
max. 100 mA
(2 x operating parameters,
2 x measuring range
feedback)
AutoCal-Function
Yes
No