63
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
Fig. 71: Third step of the manual calibration procedure
The label on the start button changes to ‘3. Step’ button.
Note
PMA1000 only: Tapping on the ‘3. Step’ button deactivates the valves
HSS1 (switching between sample/test gas) and also HSS2 (for zero
gas) or HSS3 (for span gas). This resets the trigger of the I/O5 hardware
components (connector X52).
Please tap on the ‘3. Step’ button to save the measured value.
Fig. 72: End of the manual calibration procedure
The label on the start button changes to ‘Complete’.
Note
To continue the manual calibration procedure with another test gas,
please scroll to ‘Zero gas’ or ‘Span gas’.
Manual calibration with ‘Zero gas’ or ‘Span gas’ can be repeated at
any time.
Please tap on the ‘Complete’ button to reset the status relay RS2, which is part of the I/O2
hardware components (connector X20). This tap ends the calibration mode.
After tapping on the ‘Complete’ button the display immediately goes back to the start
screen.