60
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
10 Calibration
10.1 General
There are currently two analyzer models available: PMA1000 and PMA1000L. The function-
ality of the L-model includes manual calibration. The PMA1000 features an automated cali-
bration tool (AutoCal).
To calibrate the analyzer, you need a test gas with a known oxygen concentration. During
the calibration phase the corresponding mA value to the known oxygen concentration is
applied to the mA-output.
Hazardous
Gas!
Caution, hazardous gas! Do not inhale!
10.2 M5/S1 Manual Calibration
Select your test gas and set calibration parameters
The manual calibration of the analyzer is easy and simple. Please start by selecting your
test gas. You can choose between zero gas or span gas.
Note
Please don’t forget to use the scroll bar and select ‘Zero gas’ or ‘Span
gas’. An error message will open, when the test gas is not selected.
1
3
5
4
2
7
6
Fig. 66: Manual calibration PMA1000L
1
Start button
2
Scroll bar to select test gas
3
Measuring range of calibration in brackets
4
Message box
5
Edit button
6
Test gas concentration
7
Calibration button M5