43
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
Channel Parameter description
Default
value
Selection: 2= Hardware configuration
K1
mA range 1=0-20 mA, 2=4-20 mA
2
K1
Gas flow from Kx (1...n)
1
K1
pressure reading on screen Kx (1...n) enable=0, disable=1
0
K1
flow reading on screen Kx (1...n) enable=0, disable=1
0
K1
Option negative reading ignore: 0=no, 1=yes, active
0
K1
Option pressure compensation: 0=no, 1=P_In, 2=P_Out
0
Selection: 3= Calibration
K1
zero gas [vol%]
0.000
K1
span gas [vol%]
20.960
K1
Holding time [s] of relay 2 after calibration
1
K1
Calibration: MIN range zero gas [vol%]
-2.000
K1
Calibration: MAX range zero gas [vol%]
2.000
K1
Calibration: MIN range span gas [vol%]
19.000
K1
Calibration: MAX range span gas [vol%]
24.000
K1
Calibration: MeasRange for zero gas
1
K1
Calibration: MeasRange for span gas
4
K1
AutoCal: zero gas [vol%]
0.000
K1
AutoCal: span gas [vol%]
20.960
K1
AutoCal: MIN range zero gas [vol%]
-2.000
K1
AutoCal: MAX range zero gas [vol%]
2.000
K1
AutoCal: MIN range span gas [vol%]
19.000
K1
AutoCal: MAX range span gas [vol%]
24.000
K1
AutoCal: MeasRange for zero gas
1
K1
AutoCal: MeasRange for span gas
3
Selection: 4= Measuring range selection
K1
measuring range at start
3
K1
measuring range 2 from [vol%]
0.000
K1
measuring range 2 to [vol%]
10.000
K1
measuring range 3 from [vol%]
0.000
K1
measuring range 3 to [vol%]
30.000
Selection: 5= Thresholds (Lim)
K1
range op. lim 1 lim2 +- [vol%]
0.000
Selection: 6= Sensor assignment
K1
Assignment sensor module values (No. 1-3)
1
K1
..IR bench OUT3A-HSSW output for active Kxn (0=not active)
0
K1
..IR bench A8-OUT Meas. ch2 for active Kxn (0=not active)
0
K1
..IR bench A8-OUT Meas. ch3 for active Kxn (0=not active)
0