58
PMA1000/PMA1000L | 1.00.00
www.mc-techgroup.com
9 Starting-up and operating the analyzer
9.1 General
Please be sure to observe the generally accepted engineering standards, and all of your
national and local regulations before starting up the analyzer.
Ensure that the specified voltage displayed on the product label matches the available
supply voltage before connecting the device to the supply voltage.
NOTICE
Incorrect voltage may damage the device.
The supply voltage must match the technical data displayed on the
product label.
9.2 Start-up and operation
The operation temperature of the paramagnetic measuring cell is 55 °C (131 °F). After
turning the analyzer on, the device starts to warm-up. During this warm-up phase, the
current temperature is shown on the screen. The yellow light indicates that the device is
not ready to operate yet.
An accurate measurement during the warm-up phase is not possible. After the device has
reached the operation temperature, the start screen with the measured values will be
automatically displayed on the screen.