41
Charging cancelled: check the electrodes.
Solution:
Shut down the AED7000 and check pads connector. If the
connector inadequately connects to the socket of the AED7000,
push the connector firmly into the socket.
Shut down the AED7000 and check the pads contacting with
patient’s bare chest. If the pads are poorly contacted, wipe off
any moisture and press the pads firmly on the chest of the
patient.
6.3.4
Troubleshooting of defibrillation
Problem description:
The AED7000 automatically has an internal disarm function following completion
of the analysis process. The unit broadcasts a voice prompt of “Time out, analysis
restart”.
Time out, analysis restart.
Solution:
If the operator doesn’t press the shock button in 30
seconds after the analysis is finished, the AED7000 will
automatically disarm.
Press the shock button in 30 seconds after the voice
prompts of “press flashing shock button”.
Press flashing shock button.