34
At the end of 10 seconds of CPR, the AED7000 will give a prompt to indicate that
the operator must stop CPR and not touch the patient so that the AED7000 can
restart a heart rhythm analysis and determine if a shockable rhythm condition
exists.
Stop CPR, start analysis ten seconds later.
5.9
An emergency cancellation
If any unpredictable situation occurs, the operator can use the ON/OFF button to
make an emergency cancellation.
Examples of unpredictable situation may include:
The movement of the patient during the discharging period.
The disconnection of the electrode pads during the discharging period.
Other dangerous situations.
If an unpredictable situation occurs, the operator should press the ON/OFF button
for 2 seconds to shut down the AED7000 and internally discharge all the power to
the machine