39
Caution
When cleaning the electrode pads socket of the
AED7000, do not use excess water or cleanser.
6.2.2
Disinfecting
Disinfection of the AED7000 is recommended. Use only alcohol
based solvents.
Caution
Never use the following cleansers on the AED7000:
Acetone, Ammonia cleanser, Glutaraldehyde.
6.3
Troubleshooting
As with any equipment, problems may occur during the operation of the AED7000.
Indicators and voice prompts will be given to the operator to indicate any identified
issues with the unit.
General trouble shooting tips are noted below.
Note
The operator could try to salve the problem which occurring prior
to or during the operation according to these trouble shootings
beblow
6.3.1
The troubleshooting of attaching electrode pads
Problem description:
The voice prompt of “Check the electrode” continues to broadcast when the
electrode pads have been placed on the patient and the pads connector has been
connect to the socket of the AED7000.