47
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
Medical Electrical Equipment requires special precautions regarding EMC and
must be installed and put into service according to EMC information provided in
this document.
7.6.2
Electromagnetic immunity (IEC 60601-1-2 Table 202)
The AED7000 is intended for use in the electromagnetic environment specified
below. The customer or the user of the AED7000 should ensure that it is used in
such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment–guidanc
e
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output
lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1kV
differential
mode
±2kV common
mode
Not applicable
Not applicable