13
water or other liquids
Do not use the defibrillator if the unit has been
immersed in liquid or if excessive
condensation is visible on the device.
Caution
Conductive parts should not contact other
conductive parts including the earth
2.4.3
Improper Device Performance
Warning
Properly use electrode pads
Do not attempt to warm the electrodes with a
heat source greater than 35°C (95°F). Do not
immerse or clean electrodes with alcohol or
solvents. Do not perform chest compressions
(CPR) through electrodes. These actions may
damage the electrode pads and cause the unit to
function improperly.
Warning
Use only M&B approved accessories
Do not use defibrillation pads, batteries, and
other accessories not approved by M&B. Use
of unauthorized accessories may cause the
device to operate improperly and provide false
measurements. Follow all labeling instructions
on the defibrillation pads and the battery.
Caution
Do not repeatedly charge and discharge
defibrillator in rapid succession
If a need for repetitive testing arises, wait at
least 1 minute for every third discharge to
avoid damaging equipment.
Caution
Improper maintenance can cause improper
performance