WDS
Operation, Maintenance and Installation Manual
Issue 11
Page
38
14/02/2019
LEVEL 0
LEVEL 1
SELECT
COMMENTS
SENSORS
BLOC ON/OF
Blockage
DLOK ON/OF
Door Lock
DINT ON/OF
Door Interlock
PAN1 ON/OF
Panel 1 Interlock
PAN2 ON/OF
Panel 2 Interlock
RACK ON/OF
Rack Sensor
LEVW n
Water Level
LEVS n
Steam Water Level
TCH nnn
Chamber Temperature nnn DegC
TWA nnn
Water Temperature nnn DegC
TSW nnn
Steam Water Temp nnn DegC
V5 nn
Supply Voltage 5V
V12 nn
Supply Voltage 12V
V24
Supply Voltage 24V
EXIT
DETERG
START nn
Operates Det Pump for dosage – nn mL – Press Enter to Start
EXIT
LCD
LCD TST
Press Enter to Start LCD Test. Displays all SEGs on for 3sec on the
OFF
EXIT
RUN
BP-UR
Select Cycle Mode 0
URINE
Select Cycle Mode 1
BOWL
Select Cycle Mode 2
CF1
220: Cold Fill 1
CFLUSH
240: Cold Flush
CF2
260: Cold Fill 2
CWASH
280: Cold Wash
HFD1
300: Hot Fill + Detergent
HWASH
320: Hot Wash
HF2
340: Hot Fill 2
HRINS
360: Hot Rinse
STEAM
400: Steam Heatup and Disinfect
COOL
500: Cool Down
DRY
520: Drying
START
START CYCLE- Cycle Mode (0 to 2) starting at Cycle (220 to 520)
EXIT
EXIT
SERVICE
DIAGNOS
PWRUPS
Number of Power-ups / Reset
RUN HRS
Total Running Time (Hrs)
RUN SEC
Total Running Time (Sec)
PMP HRS
Pump Run Hours (Hrs)
PMP SEC
Pump Run Hours (Sec)
CYC STA
Cycle Starts
CYC SUC
Cycles Successful
CYC FLT
Cycles Faults
EXIT
RESET
CONT TM
Enter nn==7 and then
Confirm
Continuous Test Mode
EXIT