3.2 Urinal Bottle / Bedpan – Operating Cycle
Refer to laminated WDS Instructions
1. Open door, use
or press
2. Select URINE / BEDPAN cycle by adjusting the rack position as shown
3. Close door, use the or press
4. When the door closes, to start cycle use the
or press
whilst “START” is flashing for 10 sec
Alternatively the
can be pressed to close the door and automatically start the cycle
Cycles will be shown on LCD display
1.
Flush
2.
Cold water wash
3.
Hot water wash with detergent
4.
Hot water rinse
5.
Disinfection
6.
Cool down
7.
Completed
When the cycle is completed the display will show COMPLETED. OPEN DOOR
Open door, use
or press
NOTE
: If temperature inside the chamber is too high the display will flash,
WAITING COOLING
When the temperature has dropped to a safe level the display will flash,
OPEN DOOR
Open door, use
or press
To run the same cycle repeat steps 1 to 3
If you don’t need to run another cycle, close door by pressing
NOTE:
When the cycle has started the operating cycle cannot be interrupted
In an emergency, if the cycle needs to be stopped press the EMERGENCY CYCLE STOP button
The display will show a Fault code 904 (see note below)
NOTE:
If a FAULT code is displayed on LCD display contact the Service Engineer to rectify