![Malmet ES915 Operation, Maintenance And Installation Manual Download Page 7](http://html2.mh-extra.com/html/malmet/es915/es915_operation-maintenance-and-installation-manual_3918624007.webp)
Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5
Page
6
June 23
1.2
Chemical Dosing System
The chemical dosing system ensures the correct amount of process chemicals are delivered at the correct stage during each
operating cycle. Chemical addition is controlled by the automatic cycle controller and will display a fault if the incorrect
amount of chemical is injected. The dosing system is designed for use with Malmet Chemical; other chemicals may impair
the operation of the system.
1.3
Detergent
Malmet Washer Disinfectors are designed and tested using Malmet specific detergent. To ensure intended performance
and trouble-free operation use only Malmet branded detergent, failure to use correct detergent may void warranty.
The detergent system delivers 30ml of concentrated detergent in the initial wash stage.
Changing the Detergent Bottle
WARNING!
SAFETY GLOVES AND GOGGLES MUST BE WORN WHEN CHANGING
DETERGENT AND DISPOSING OF EMPTY CONTAINERS
Current Safety Data Sheet for Malmet detergent is available in .PDF format from Malmet’s website.
Only use Malmet approved detergent (See technical data for detergent details)
1.
Pull latch on detergent door and open.
2.
Unscrew cap and pull out with suction hose (let hose hang in detergent chamber).
3.
Remove empty bottle and replace with full bottle.
Note: Leave cap on new bottle until in position.
4.
Remove cap on new bottle and fit existing hose and cap
,
make sure suction hose is at bottom of bottle
5.
Close detergent door.
6.
Restart device operation as normal.
Disposal of Empty Detergent Bottles
1.
When handling empty containers treat as though container is full and wear appropriate PPE as per MSD requirements.
2.
Triple rinse container and puncture base to render unusable.
3.
Dispose of containers as per facility chemical waste handling procedures, and/or local government requirements.