June 23
•
Follow your internal manual handling guidelines to manoeuvre the device off the pallet. The device can then be placed
into position by fork lift or hand pallet truck.
Disposal of Packaging
•
Please dispose of packaging as per facility procedures or local government requirements.
2.1.1 Positioning the ES-D
Model
Placement
Access
Required
Device Dimensions
Height (mm)
Width (mm)
Depth (mm)
ES915 models
ES935 models
Freestanding
Both Sides
1305
595
660
Please allow sufficient room for servicing purposes. Recommended space requirements 200mm on either side and
150mm at the rear of the device.
New Buildings
•
Service connections are normally pre-placed after planning and consultation with all interested parties. Installation is by
connection to the services provided.
•
As the soil line (sewerage outlet) is the least flexible of all the connections, this usually influences the decision as to
where to place the Bedpan / Urinal Bottle Washer Disinfector. If an existing soil line can be utilised this will represent a
cost saving.
•
The Bedpan / Urinal Bottle Washer Disinfector is supplied with either a ‘S’ or ‘P’ Trap as nominated by the Purchaser.
The ‘S’ Trap connects through the floor and the ‘P’ Trap connects through the back wall. The trap section is easily
removed if the wrong trap has been ordered. Refer to Diagram C2 for trap connections.
•
Potential electromagnetic or other interference between other EQUIPMENT and other devices can possibly affect the
Infra-red hands-free operation sensor. It is advisable to check all the equipment and devices in the intended
installation area that have infra-red operation.
Electromagnetic interference can be prevented by installing the device
in non-patient areas of the facility (or similar).