![Malmet ES915 Operation, Maintenance And Installation Manual Download Page 27](http://html2.mh-extra.com/html/malmet/es915/es915_operation-maintenance-and-installation-manual_3918624027.webp)
Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5
Page
26
June 23
4.1
Display Board
Green LED Display
Door open
OPEN
Door closed
Shut
Steam generator fill
Fill
Cycle ready
rdY
Cycle started
rinS
Detergent
dEt
Pause
PAUS
Disinfection cycle started
(flashing)
90˚C
Disinfection temperature reached
‘actual’ temp ˚C
Cool down final rinse
cool
End of cycle
End
Chamber above safe opening temperature
Hot
Yellow LED illuminates on control panel
Overflow
2 flashing green LED in activation area
(8 seconds)
Cycle can be started
4.2
Release of Processed items
On completion of the cycle, it is recommended that all processed items are unloaded and visually inspected as clean as per
the requirements AS/NZS 4187. It is recommended all items are removed on completion of the cycle and stored in a
designated area to prevent any risk of contamination or mixing with unprocessed items.
If the processing status of a load or item is unknown or in question, it should be treated as unprocessed and not
released.