Operation Qualification
Performed immediately after installation, relocation, change of a service to the device or after repair prior to device being
put back in service in accordance with applicable Standards.
Performance Qualification
Performed immediately after IQ or OQ for new installs or relocated equipment, or change of a service to the device or
repair that might adversely impact the quality of the RMD.
5.10 Requalification
It is recommended requalification of the process is performed
annually in accordance with the requirements of
AS/NZS4187 and ISO 15883 applicable parts
5.11 Device faults
In the event of a fault or malfunction with the device during operation, abort the cycle by pressing the Power button on
the control LCD. Once the device is off, isolate the mains power supply and water services. Notify maintenance or service
provider of the fault and ensure device is not used until the fault is rectified.
5.12 Fault Indication
The device is controlled by a micro-processor. The processor has fault detection capability and indicates faults by code
on the digital display; Diagram B1 (below) shows an example of fault code displayed.
Diagram B1
(Fault 01 indicated)
Please see table in 5.13 Fault List for a full list of fault codes along with their description and possible cause.
Note: If a fault is displayed on the LED display, turn device off and turn it back on, press START to try to complete a cycle.
If a fault persists contact a Service Technician. Record the fault code that has appeared for the Service Technician.
Table B1 indicates these faults. For a more detailed description of these fault conditions see Appendix A
–
Faults.