Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5
Page
25
June 23
19.
After the disinfection cycle is complete the cooling solenoid will turn on for 15 seconds. This will cool down the
surface temperature of the utensils and make them safe to handle.
The display shows cool
.
This water is direct from the mains.
20.
At the end of the cool down rinse,
the display shows End
. This will be displayed until the door is re-opened. If the
chamber temperature is above the safe opening temperature the
display shows “HOT” and access is restricted until
temperature has reduced to a safe level.
21.
If the Bedpan / Urinal Bottle Washer Disinfector is not used, the steam generator temperature will be kept at 85°C-
90°C for a period of 15 minutes. After 15 minutes, if the Bedpan / Urinal Bottle Washer Disinfector is still not used it
will enter into idle mode and the elements will only come on at 50°C and rise to the temperature of 90°C repeatedly
until the cycle is started again.