Malmet Bedpan / Urinal Bottle Washer Disinfector (ES-D)
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 5
Page
14
June 23
2.5
Electrical
These installation guidelines must be followed to ensure the device will operate as intended and must be carried out by
a qualified, licensed electrical tradesperson. The device must be installed and serviced to national wiring rules AS/NZS
3000
.
Model:
ES915 models
240V 50Hz
1 Ph
20Amp
ES935 models
240V 50Hz
1 Ph
10Amp
Note: Mains power connection
Overcurrent Protection device: A circuit breaker or fuse must be installed in the building installation for the device.
10 Amp Devices
–
To be hard wired to a lockout ISO switch or Circuit breaker.
Switch must be wired to 10A max circuit breaker at main switch board
20 Amp Devices
–
To be hard wired to a lockout ISO switch or Circuit breaker.
Switch must be wired to 20A circuit breaker at main switch board
•
An ISO switch or circuit breaker must be included in the installation. (Not supplied by Malmet)
•
Mains power ISO switch must be suitably located and easily reached, approximately 1500mm above floor level adjacent to
device. The 1.7m mains power lead exits the device approximately 1200mm above floor level on the right-hand side of the
device.
•
It must be marked as the disconnecting device for the equipment.
•
It must have contact disconnection of all poles to provide full disconnection.
•
Ensure isolator is suitably placed so device will not obstruct safe access to the isolator once installed.
•
If the supply lead is damaged, it shall be replaced by Malmet; its authorised service agent or similarly qualified person in
order to avoid hazard.
Mains power ISO switch or circuit breaker must be in an accessible position so device can be isolated from mains power
during service.
All devices must be earthed.
The maximum permissible system impedance Zsys:
Z = 0.220 Ohm + j 0.137 Ohm (0.220 Ohm + 437 μH)