<5> Checking the movement range
After mounting the scale part and slider part, be sure to move the
machine over the overall length of the scale to check that the
machine moves within the measuring length.
Remember that the scale unit may be damaged if the machine
moves beyond the movable range of the scale unit (measuring
leeway).
Secure the cable clamp using a M4
×
5
cable clamp screw (accessory).
Fig. 3-14
<6> Mounting the headcable
Fix the head cable with cable clamps so that they will not
interfere with moving parts of the machine.
Note
The head cable has an integrated design with the scale unit, and
so mount the scale unit while being careful not to put pressure on
the cable. Take care not to pull the head cable forcefully or bend
it repeatedly or the cable may break.
Fig. 3-13 Removal of slider holder
Slider holder
Remove the four M2.6 screws.
Adhere the screening label
(accessory).
<4> Removal of the slider holder
Remove the four M2.6 mounting screws for the slider holder and
then remove the slider holder itself.
•
Be sure to adhere the screening label (accessory) to cover the
screw holes after removing the slider holder. If these holes are
not plugged, chips, cutting fluid or dust may enter and the
accuracy of the scale may be deteriorated. Use a ball-point pen
to enter the mounting date and name of the person responsible
on the label.
Note
Keep slider holder and four M2.6 screws after the installation.