22
(E)
MG10A-P1 / MG10A-P2
5-4-2. Details of Command Settings
ASCII code is used for all the commands. A command is
terminated by “
CR
” or “
CR
” + “
LF
”. (“
CR
”: carriage return; “
LF
”:
line feed)
• Select using communication setting switch No. 2. on main
unit.
OFF
:
“
CR
” + “
LF
” (factory setting)
ON :
“
CR
”
• On the pages to follow, “
c
” and “
d
” are used to indicate the
value to be set.
• When inputting a setting, a numerical value inside the
assigned range must be selected: otherwise, the value will not
be set.
• If no polarity (+ or −) is input, it will be construed as +.
[Data which is input ahead of the commands]
Data
Selection range
Symbol used in
descriptions
Unit number
0 to F or
*
a
Module number
0 to F or
*
b
• 0 to F are hexadecimal numbers which are equivalent to 0 to
15 under the decimal system.
•
*
is used when the settings are to be established for all units or
all modules.
Setup commands
Command Description
Write/Read
SETUP
Start parameter setup.
Command
SETUP
Starts parameter setup.
MODE=/MODE=?
Set the measuring mode.
Command/
ab
MODE=
c
Return format
c
= 0 : Current value (REAL) mode
1 : Maximum value (MAX) mode
2 : Minimum value (MIN) mode
3 : Peak-to-peak value (P-P) mode
P=/P=?
Set the preset value.
Command/
ab
P=
c
Return format
c
= The value diff ers depending on the
input resolution.
Resolution Setting
2-digit integer
+
decimal point
+
4 decimal places
0.1 μm −99.9999 to
+
99.9999
0.5 μm −99.9995 to
+
99.9995
(last digit must always be a 5 or 0)
3-digit integer
+
decimal point
+
3 decimal places
1 μm
−999.999 to
+
999.999
5 μm
−999.995 to
+
999.995
(last digit must always be a 5 or 0)
4-digit integer
+
decimal point
+
2 decimal places
10 μm −9999.99 to +9999.99
Summary of Contents for MG10A-P1
Page 2: ......
Page 3: ...J 1 MG10A P1 MG10A P2...
Page 4: ...2 J MG10A P1 MG10A P2 DC AC...
Page 6: ...ii J MG10A P1 MG10A P2 Microsoft Windows Windows 7 34 8 35 8 1 35 8 2 36 8 3 36 9 37...
Page 9: ...J 3 MG10A P1 MG10A P2 1 3 BCD DG DT DL DK PLC 16 RS 232C...
Page 10: ...4 J MG10A P1 MG10A P2 2 6 7 4 5 8 2 9 10 11 12 13 14 1 3...
Page 12: ...6 J MG10A P1 MG10A P2 3 3 1 16 1 2...
Page 13: ...J 7 MG10A P1 MG10A P2 2 3 1 1 4 3 4 3 2 I F I F 1 1...
Page 14: ...8 J MG10A P1 MG10A P2 I F I F 2 1 3 3...
Page 15: ...J 9 MG10A P1 MG10A P2 DIN DIN DIN 2 3 3 DIN DIN DIN 35 mm DIN 1...
Page 16: ...10 J MG10A P1 MG10A P2 3 5 3 I O 3 GND 2 Vin DC 12 V 24 V 1 FG 3 1 2 3 4 1...
Page 17: ...J 11 MG10A P1 MG10A P2 3 6 LZ61 CE08 1 3 7 RS 232C DZ252 RS 232C RS 232C I F 1...
Page 19: ...J 13 MG10A P1 MG10A P2 4 2 3 DK 1 1 mm 1 mm 2 3...
Page 20: ...14 J MG10A P1 MG10A P2 4 2 5 P P RS 232C SETUP P CLOSE 1 mm RS 232C 1 4 2 4 0 4...
Page 21: ...J 15 MG10A P1 MG10A P2 4 2 6 RS 232C MG30 B BCD 4 RS 232 NG U GO G GO G GO G NG L...
Page 22: ...16 J MG10A P1 MG10A P2 ON P P OFF 4 2 7...
Page 24: ...18 J MG10A P1 MG10A P2 5 2 5 2 1 0 F 1...
Page 31: ...J 25 MG10A P1 MG10A P2 6 RS 232C I O I F I F I F 4 2 3 OFF RS 232C I O...
Page 39: ...J 33 MG10A P1 MG10A P2 6 2 3 MIN 25 ms MIN 1 ms RS RS MIN 2 ms MAX 2 ms MIN 2 ms RS 232C...
Page 43: ...J 37 MG10A P1 MG10A P2 12 V 24 V RS 232C MG10A MG10A MG10A 9 MG10A...
Page 44: ...38 J MG10A P1 MG10A P2...
Page 52: ...4 E MG10A P1 MG10A P2 6 7 4 5 8 2 9 10 11 12 13 14 1 3 2 Name and Function of Each Part...
Page 86: ...38 E MG10A P1 MG10A P2...
Page 94: ...4 G MG10A P1 MG10A P2 6 7 4 5 8 2 9 10 11 12 13 14 1 3 2 Teilebezeichnungen und Funktionen...
Page 129: ......
Page 131: ...MG10A P1 MG10A P2 1...