XENON LAMP REPLACEMENT

To replace the lamp:

1. Orient lamp to face forward. See Figure 8.  Lower lamp into the CLX with lamp against fan.  See figure 9.

The lamp receptacles must engage with lamp base plugs.  Press firmly into place to seat lamp.

Note: Do not touch lamp glass window.

2. Snap the lamp retainer clips, located on each side of the lamp, onto the top of the lamp.  The lamp is now in a

locked position.  See Figure 6.

3. Fully slide the top cover of the CLX forward.  Install the retaining screw (Figure 3) using a screwdriver.

4. Connect power cord to the back of the CLX.  Set main power on/off switch to the “on” position then press on / 

standby switch to turn the unit “on”.

5. Adjust attenuator ring to verify that the unit emits light.

Figure 9

Figure 8

Forward

Lamp module

Lamp disposal:

No disposal restrictions are known for the Xenon lamp.  User should check and follow any applicable federal, 
state and/or local regulations.

To dispose of Luxtec Xenon lamps, simply break the fill tab to release any remaining gas and then dispose of the 
lamp with no environmental concerns.  Luxtec lamps are constructed entirely of metal, sapphire and ceramic.  
No organic (carbon-based) materials, mercury, rare-earth elements, or any other materials with disposal 
problems are used in the construction of our lamps.  The Xenon fill gas is an inert gas and is non-toxic.

- 6  -

Summary of Contents for 00-CLX

Page 1: ...User Manual for CLX Xenon Light Source ENGLISH...

Page 2: ...eshooting 9 Maintenance and Cleaning 10 Specifications 10 Electromagnetic Compatibility EMC User Information 11 14 Product Warranty 15 Repair and Return 15 TABLE OF SYMBOLS Found on Light Source Atten...

Page 3: ...and the light cable can cause eye damage if viewed directly for more than momentary periods of time Do not look directly into the light source or the cable Do not point the light cable at another pers...

Page 4: ...The CLX light source comes with a hospital grade power cord Please verify that both the light source and the power cord have been received undamaged All fiber optic light guides should be properly cle...

Page 5: ...t occur when the standby switch is used to control the lamp Note that if the light source is left in standby with power applied with the main power on off switch in the on position and the on standby...

Page 6: ...nsure against failure Figure 7 Figure 6 Retaining clips locked Replace lamp when illumination is no longer satisfactory or sufficient for use To remove the lamp 1 Make sure the main power on off switc...

Page 7: ...main power on off switch to the on position then press on standby switch to turn the unit on 5 Adjust attenuator ring to verify that the unit emits light Figure 9 Figure 8 Forward Lamp module Lamp dis...

Page 8: ...dle screw positioned at the top of the handle slide the handle onto the upper column so there is about 5 inches 13 cm of space between the handle and the top of the column Tighten handle screw to hold...

Page 9: ...e 850358 User Manual for the CLX Adapter Options 401193 Storz adapter Lamp 00 1187 CLX Lamp replacement module Power Cords 601949 AUSTL6M Australian Power Cord 6m 601949 EUR6M European Power Cord 6m 6...

Page 10: ...air the unit if the lamp fails in use Turn off the unit and allow it to cool for at least 15 minutes then try to restart the unit If the lamp still fails to illuminate it may need to be replaced If yo...

Page 11: ...nominal AC Power Leakage Leakage current to chassis with ground wire intact less than 100 microamps Leakage current to chassis with ground wire interrupted less than 500 microamps Classification Type...

Page 12: ...ironment of use to permit the CLX Light Source to perform its intended use without disturbing other Equipment and Systems or non medical electrical equipment Guidance and Manufacturer s Declaration Em...

Page 13: ...commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment Volta...

Page 14: ...vironment specified below The customer or user of the CLX Light Source should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environm...

Page 15: ...or user of the CLX Light Source can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the CLX Light Source as reco...

Page 16: ...uxtec subjected to unusual physical or electrical stress altered with non Luxtec parts or damaged during shipment back to Luxtec This warranty is non transferable unless authorized in writing by Luxte...

Page 17: ...the Essential Requirements specified in Annex 1 of the Medical Device Directive MDD 99 Hartwell Street West Boylston MA 01583 USA Telephone 800 325 8966 USA only or 1 508 835 9700 Facsimile 1 508 835...

Reviews: