GENERAL WARNINGS
FIRE HAZARD:
Do Not
drape or cover the light source while it is operating
EXPLOSION HAZARD:
Do Not
use in the presence of flammable anesthetics.
ELECTRIC SHOCK HAZARD:
Do Not Open
the light source if it is not functioning properly or if damage
is apparent (such as case damage or loose screws). Refer to the Repair and Return Section of this
Manual.
•
The User of this equipment should carefully read the Users Manual and be familiar with the set-up,
use, and care of the light source before using it. The instructions in the Users Manual should be
followed carefully with special attention to warnings and control features on the light source. The
Users Manual should be readily available to users and other appropriate personnel.
•
Before each use ensure that the light source is not damaged and is operating properly. Ensure that
cooling vents and fans are free of obstruction.
•
The light source generates high intensity light that can heat objects in its path. This light can cause
flammable materials to burn or cause burns to skin. Prevent flammable materials from coming into
close proximity with the output of the light source and light cable.
•
The light emitted from the light source and the light cable can cause eye damage if viewed directly for
more than momentary periods of time. Do not look directly into the light source or the cable.
Do not point the light cable at another person’s eyes. For this reason this product must not be used
in ocular surgery or in any surgical procedure requiring direct illumination of the eye.
•
Avoid extreme heat or jarring of the light source to maintain the integrity of the Xenon lamp.
Refer to the Xenon Lamp Replacement Section of this Manual when replacing the lamp or removing
it for any reason.
•
Follow the instructions of other manufacturers equipment when using them in conjunction with this
light source.
•
This product must be removed from the surgical field prior to use of a defibrillator.
•
Before any service, disconnect the power cord from the main outlet.
Medical Electrical Equipment such as the CLX Light Source require special precautions regarding
electromagnetic compatibility or EMC. These precautions are defined on pages 11 thru 14 of this Users
Manual. The CLX must be installed and put into service according to the EMC information provided.
The leakage current of medical electrical devices can be additive. When using the CLX in combination
with other medical electrical devices, such as endoscopes and/or endoscopically-used accessories, the
leakage current should be verified to ensure it continues to meet requirements.
- 2 -