Electromagnetic Compatibility (EMC) User Information

 

Guidance and Manufacturer’s Declaration—Immunity All Equipment and Systems 

The 

CLX Light Source

 

is intended for use in the electromagnetic environment specified below. The customer 

or user of the 

CLX Light Source

 

 should assure that it is used in such an environment. 

Immunity Test 

IEC 60601 Test 

Level 

Compliance Level 

Electromagnetic Environment - Guidance 

Electrostatic 
Discharge 
(ESD) 
IEC 61000-4-2 

±6kV contact 
 
±8kV air 

±6kV contact 
 
±8kV air 
 

Floors should be wood, concrete or ceramic 
tile. If floors are synthetic, the relative humidity 
should be at least 30%

.

 

Electrical Fast 
Transient/burst 
IEC 61000-4-4 

±2kV on AC Mains   ±2kV on AC Mains  

Mains power quality should be that of a typical 
commercial or hospital environment.   
 
 

Surge  
IEC 61000-4-5 

±1kV Differential 
±2kV Common 

±1kV Differential 
±2kV Common 

 Mains power quality should be that of a typical 
commercial or hospital environment.   
 

Voltage dips, 
short 
interruptions 
and voltage 
variations on 
power supply 
input lines 
IEC 61000-4-11 

>95% Dip for  
0.5 Cycle 
 
60% Dip for  
5 Cycles 
 
30% Dip for  
25 Cycles 
 
>95% Dip for 
5 Seconds 

>95% Dip for  
0.5 Cycle 
 
60% Dip for  
5 Cycles 
 
30% Dip for  
25 Cycles 
 
>95% Dip for 
5 Seconds 

Mains power quality should be that of a typical 
commercial or hospital environment. If the user 
of the 

CLX Light Source

 requires continued 

operation during power mains interruptions, it 
is recommended that the 

CLX Light Source

 

be powered from an uninterruptible power 
supply or battery. 

Power 
Frequency 
50/60Hz 
Magnetic Field 
IEC 61000-4-8 

3A/m 3A/m 

 

Power frequency magnetic fields should be 
that of a typical location in a typical commercial 
or hospital environment. 

 

 
 

- 12 -

Summary of Contents for 00-CLX

Page 1: ...User Manual for CLX Xenon Light Source ENGLISH...

Page 2: ...eshooting 9 Maintenance and Cleaning 10 Specifications 10 Electromagnetic Compatibility EMC User Information 11 14 Product Warranty 15 Repair and Return 15 TABLE OF SYMBOLS Found on Light Source Atten...

Page 3: ...and the light cable can cause eye damage if viewed directly for more than momentary periods of time Do not look directly into the light source or the cable Do not point the light cable at another pers...

Page 4: ...The CLX light source comes with a hospital grade power cord Please verify that both the light source and the power cord have been received undamaged All fiber optic light guides should be properly cle...

Page 5: ...t occur when the standby switch is used to control the lamp Note that if the light source is left in standby with power applied with the main power on off switch in the on position and the on standby...

Page 6: ...nsure against failure Figure 7 Figure 6 Retaining clips locked Replace lamp when illumination is no longer satisfactory or sufficient for use To remove the lamp 1 Make sure the main power on off switc...

Page 7: ...main power on off switch to the on position then press on standby switch to turn the unit on 5 Adjust attenuator ring to verify that the unit emits light Figure 9 Figure 8 Forward Lamp module Lamp dis...

Page 8: ...dle screw positioned at the top of the handle slide the handle onto the upper column so there is about 5 inches 13 cm of space between the handle and the top of the column Tighten handle screw to hold...

Page 9: ...e 850358 User Manual for the CLX Adapter Options 401193 Storz adapter Lamp 00 1187 CLX Lamp replacement module Power Cords 601949 AUSTL6M Australian Power Cord 6m 601949 EUR6M European Power Cord 6m 6...

Page 10: ...air the unit if the lamp fails in use Turn off the unit and allow it to cool for at least 15 minutes then try to restart the unit If the lamp still fails to illuminate it may need to be replaced If yo...

Page 11: ...nominal AC Power Leakage Leakage current to chassis with ground wire intact less than 100 microamps Leakage current to chassis with ground wire interrupted less than 500 microamps Classification Type...

Page 12: ...ironment of use to permit the CLX Light Source to perform its intended use without disturbing other Equipment and Systems or non medical electrical equipment Guidance and Manufacturer s Declaration Em...

Page 13: ...commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment Volta...

Page 14: ...vironment specified below The customer or user of the CLX Light Source should ensure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environm...

Page 15: ...or user of the CLX Light Source can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the CLX Light Source as reco...

Page 16: ...uxtec subjected to unusual physical or electrical stress altered with non Luxtec parts or damaged during shipment back to Luxtec This warranty is non transferable unless authorized in writing by Luxte...

Page 17: ...the Essential Requirements specified in Annex 1 of the Medical Device Directive MDD 99 Hartwell Street West Boylston MA 01583 USA Telephone 800 325 8966 USA only or 1 508 835 9700 Facsimile 1 508 835...

Reviews: