Electromagnetic Compatibility (EMC) User Information
Please note: The CLX light source has been successfully tested to IEC 60601-2-25. The
CLX was in the presence of an electrical-surgical generator, ESG, generating a field
strength of between 18 and 23 V/m. During the testing the energized probes of the
ESG were at a distance of 6" from the CLX and the ESG was energized in the cut and
coagulate modes. The functioning of the CLX was not effected by exposure to this
energy. The results of this testing are available from Luxtec.
Guidance and Manufacturer’s Declaration –
Emissions Equipment and Systems that are NOT Life-Supporting
The
CLX Light Source
is intended for use in the electromagnetic environment specified below. The
customer or user of the
CLX Light Source
should ensure that it is used in such an environment.
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to
80 MHz
V1 = 3 Vrms Portable and mobile RF communications equipment should
be separated from the
CLX Light Source
by no less
than the
recommended separation distances calculated/listed below:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
E1 = 3V/m
P
D
17
.
1
=
80 to 800 MHz
P
D
33
.
2
=
800 MHz to 2.5 GHz
Where P is the maximum power rating in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as determined by an
electromagnetic site survey, should be less that the
compliance levels (V1 and E1).
Interference may occur in the vicinity of equipment containing
a transmitter.
P
D
17
.
1
=