Electromagnetic Compatibility (EMC) User Information
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the CLX Light Source
Equipment and Systems that are NOT Life-Supporting
The
CLX Light Source
is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The
customer or user of the
CLX Light Source
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF Communications Equipment and the
CLX Light Source
as recommended below, according to the maximum
output power of the communications equipment.
Recommended Separation Distances for the CLX Light Source
(meters)
Maximum
Output
Power
(Watts)
150 kHz to 80
MHz
P
d
17
.
1
=
80 to 800MHz
P
d
17
.
1
=
800 MHz to 2.5 GHz
P
d
33
.
2
=
0.01 0.12
0.12
.023
0.1 0.37
0.37
0.74
1 1.17 1.17
2.33
10 3.69
3.69
7.38
100 11.67
11.67
23.33
- 14 -