21.3 I/O Ports (including RS-232)
21.3.1
The S2000 has 5 different I/O ports; 4 located on the back of the unit. The 1st
port is a simple 2-wire, audio style jack that can be connected to a foot pedal
(supplied with each unit) or any other electro-mechanical triggering device. This
is an exposure trigger input used to start an exposure. The 2nd port is a RS-
232 port (9 pin) to be connected to a PC and is described in more detail in
section 19. The 3rd and 4th ports (15 pin DE style) are for PLC interfacing. A
5thport is a RS-232 port located on the side of the front panel which is used to
connect to the R2000 Radiometer for calibration of the S2000.
21.4 USB Interface
21.4.1
A USB interface, an optional feature, can be provided through the existing RS-
232 port via an external RS-232 to USB converter device. A RS-232 to USB
converter device such as part number RS232B by FTDI (Future Technology
Devices International Inc.) or equivalent can be used.
21.5 Environmental Conditions
Operating Conditions
Ambient Temperature:
15ºC to 40ºC
Altitude:
2000m max.
Atmospheric Pressure:
700 to 1060 hPa
Relative Humidity:
15% to 95% (non-condensing)
Installation Category:
II
Pollution Degree:
2
Transport and Storage Conditions
Temperature:
-40 to +70ºC
Relative Humidity:
10% to 100%
Atmospheric Pressure:
500 to 1060 hPa