User Guide
- 54
–
18.1.6
Download: The download button transfers the step cure process into the S2000
unit. When programmed with Step Cure, initiating a timed exposure on the
S2000 unit will perform the step cure operation. In Timer or Hours mode,
pressing the start/stop button will run the step cure cycle. In Level Mode,
pressing the start/stop button will manually open the shutter until the start/stop
button is pressed again to close the shutter & the stepcure function will not be
activated.
18.2 Clear StepCure:
18.2.1
To clear the step cure from the S2000, hold the up button while pressing the
down button twice. Calibration will disable current step cure automatically
therefore Step cure has to be updated when unit calibration status changes.
Downloaded Step cure is also disabled if step cure is run directly from the GUI.
18.2.2
The clear step cure button in GUI will remove the step cure process from the
S2000 unit connected. If no step cure process is currently in the S2000 unit,
this command will have no effect.
18.2.3
Enable Password: Enter the password of
“54321” to enable the modification of
the step cure settings. The password does not need to be entered to initiate a
step cure from the control panel software.
Note:
the download function only downloads one step cure cycle.