User Guide
- 17
–
Warning
UV light will be emitted from the light guide. The light could severely damage the
cornea and retina of the eye if the light is observed directly. Eye shielding must be
used at all times as well as clothing to protect exposed skin.
8.2.2
Press the start /stop button to open the shutter. The open indicator will
illuminate. The display will show the actual irradiance output (as opposed to the
Set Point).
8.2.3
Press the up button to increase the light output intensity or press the down
button to decrease the light output intensity. For precise adjustments press the
button and release it within 1-4 sec.
8.2.4
For coarse adjustments, press and hold the up or down button for greater than
5 seconds to rapidly increase or decrease the light output.
8.2.5
Press the start /stop button to close the shutter. The open indicator will turn off.
8.2.6
To check the percent of iris opening while in the level mode; press and hold the
Mode
button. The LED display will indicate “XXX” percentage of iris opening.
Note:
While in Level Mode, pressing the start/stop button will manually open
the shutter until the start/stop button is pressed again to close the
shutter. The timer function will not be activated.
Note:
While in Absolute Mode, and while the shutter is closed, the displayed
irradiance value is the user defined Set Point. Whenever the shutter
is opened the S2000 will immediately attempt to adjust the optical
output to /-2% of the Set Point. The actual output irradiance
will be displayed at this point and will vary slightly (to /-5% or
200mW/cm2, whichever is greater) from exposure to exposure.
Note:
The shutter will not open if the light guide is not inserted properly.
Note:
OmniCure® S2000 Minimum Adjustable Irradiance Level
:
0.5W/cm²