User Guide
- 10
–
4.1.7
Locate power connector with two leads and connect it to its mate located on
the side of the lamp-housing wall.
4.1.8
Ensure the lamp anode cooling fin (bar) and lamp power lead, at the front of
the lamp, are centered within the lamp holder cut out. Rotate the lamp as
required.
4.1.9
Replace the lamp housing side panel and tighten the fastening screw.
I
NTELLI
-L
AMP
CONNECTOR
P
OWER
C
ONNECTOR
CENTER BAR AND LEAD WITHIN
THE CUTOUT
.
Note 1
: if the lamp housing panel is not secured completely the lamp will
not strike and the “
bulb
” message will display when power is
turned on to the unit.
Note 2
: if the wrong type of lamp has been installed in your unit the
message “
E1
” will appear on the LED display. Refer to section 20
for troubleshooting information.