User Guide
- 32
–
15. E3
The “E3” message appears on the display when the lamp temperature monitoring
has turned off the lamp due to excessive heat. Turn off power as per section 6 and
verify the cooling exhaust vent is not blocked. Restart the system as per section 6. If
error re-occurs the unit should be serviced.
16. E4
The “E4” message appears on the display when an internal error has occurred.
Turn off power and restart as per section 6. If error re-occurs the unit should be
serviced.
17. E5
The “E5” message appears on the display when the communication to the
intellilamp has failed. Turn off power as per section 6, open the lamp access door
and verify all connections are secure. Restart the system as per section 6. If error
re-occurs the unit should be serviced.
18. E6
The “E6” message appears on the display when an internal error has occurred.
Turn off power as per section 6, disconnect foot pedal, check for control buttons
stuck down and restart after 5 minutes. Restart the system as per section 6, if error
re-occurs the unit should be serviced.
18.
“EFI”
The “EFI” message appears on the display when an exposure failure (alarm
condition) occurs.