17.12 Alarm Function
The following conditions will cause the unit to indicate an alarm condition:
1. Lamp fails to strike (bulb)
2. Lamp access door not correctly installed (bulb)
3. Intelli-lamp sensor/ connection failure (bulb)
4. Shutter position failure (SFI)
5. Exposure failure (EFI)
17.12.1
The alarm condition is indicated by a repeating audible beep from the S2000
unit. The Unit Status indicator on the GUI will display A
larm “On”.
17.12.2
To clear the alarm signal, click on the “Clear Alarm” button. The audible alarm
on the unit will cease and the Unit Status indicator will display Alarm “Off”.
17.12.3
To clear the alarm condition press the Start/ Stop button on the front panel. The
fault which caused the original alarm condition must be corrected prior to
resuming normal operation.
17.13 Calibration Status
17.13.1
T
he S2000’s calibration interval is valid for 112 hours of elapsed lamp on time
as measured from the time the unit is calibrated with a R2000 Radiometer. The
remaining calibration time is displayed in the “Calibration In” text box in hours.
To retrieve the current remaining time, click on the GET button beside the
Calibration In text box. To turn off the calibration, click on the “Clear
Ca
libration” button. This will clear the text box, and change the unit mode to
Relative (un-calibrated) mode. To recalibrate the unit it will be necessary to
reconnect the unit to a R2000 Radiometer and repeat the calibration
procedure. Refer to section 7 for details pertaining to calibration of the Series
2000.
Note:
Performing a calibration using the RS-232 radiometer interface does NOT
require disabling the GUI.