User Guide
- 40
–
23. Get SERIES 2000 actual irradiance reading
Command to S2000;
GIM
Response from S2000;
n
(n represents a float number in Watts per
square cm)
If the unit is in LEVEL mode, “Need CAL” will be the response.
Example:
If “15.66XX\r” is responded, the S2000 irradiance reading
will be 15.66 W/cm2.
24. Get SERIES 2000 maximum irradiance output
Command to S2000;
GMP
Response from S2000;
n
(n represents a float number in Watts per
square cm)
If the unit is in LEVEL mode, “Need CAL” will be the response.
Example:
If “35.66XX\r” is responded, the S2000 maximum
irradiance output will be 35.66 W/cm2.
25. Set iris Level
Command to Series 2000;
SILn
Note
: 0 < n ≤100
n must be a integer number
Example:
If “SIL20XX\r” is set, the S2000 iris Level will be 20%.
Response from S2000;
Received
– if the value n is valid
Invalid
– if the value n is out of range
Done CAL
– the S2000 has been calibrated, so the value
can not be set.
EXP
– the S2000 is running a timed exposure, so the
value cannot be set.