User Guide
- 41
–
26. Set exposure time
Command to S2000;
STMn
Note:
2 ≤ n ≤ 9999
n must be a integer number
Example:
If “STM10XX\r” is set, the S2000 exposure time will be 1
second
Response from S2000;
Received
– if the value n is valid
Invalid
– if the value n is out of range
EXP
– the S2000 is running a timed exposure, so the
value cannot be set.
27. Set LEVEL set point
Command to S2000;
SPWn
Note
: n ≥ .1 (5mm LG)
n ≥ .25 (8mm LG) and n is a float number
Example:
If “SPW2.55XX\r” is set, the SERIES 2000 LEVEL set
point will be 2.55 W.
Response from S2000;
Received
– if the value n is valid
Invalid
– if the value n is out of range
Need CAL
– the S2000 need to be calibrated before the
value can be set.
EXP
- the S2000 is running a timed exposure, so the value
cannot be set.