Appendix G
- 246 -
MTD75 Operation Guide
Appe
ndix G
Internal Error Messages Continued
E016 UNIT
Battery
Low
The backup battery voltage is low, under normal circumstances the unit recharges
the battery. Try switching the unit off and then on again. If the message still
persists, consult with your supplier.
E017 UNIT
Water
Input
The System water flow has stopped or that there is a problem with the water
meter.
E018 UNIT
Low
Flow
The flow switch has been activated. Any relay that has been configured will be
inhibited.
E019
UNIT
SD Card Full
The SD card has become full. To continue use, either replace with a blank SD
card or remove existing files from the SD Card.
Input Card Errors
E030
Redox
Input Card Checksum Error
E080
ECS
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as the input card may require to be returned for repair.
E130
pH
E031 Redox Setup
Checksum
Error
E081 ECS
The current input cards configuration has for some reason become corrupted. Try
switching the unit off and then on again. If the message persists reset the input
card from the Reset Unit Settings option in the Save/Restore menu or consult with
your supplier, as the input card may require to be returned for repair.
E131 pH
E032
Redox
Store A Checksum Error
E082 ECS
The data in Store A has become corrupted. Check the input card’s setup. Then
save the setup back to Store A in the Save/Restore menu.
E132 pH
E033
Redox
Store B Checksum Error
E083 ECS
The data in Store B has become corrupted. Check the input card’s setup. Then save
the setup back to Store B in the Save/Restore menu.
E133 pH
E034
Redox
Factory Cal Checksum Error
E084 ECS
Try switching the unit off and then on again. If the message persists, consult with
your supplier, as the input card may require to be returned for repair.
E134 pH
E035
Redox
User Cal Checksum Error
E085 ECS
The Input Card’s User Cal has for some reason become corrupted. Try switching
the unit off and then on again. If the message persists Reset the input card from
the Reset User Calibration option in the Calibration menu or consult with your
supplier, as the input card may require to be returned for repair.
E135 pH
E086
ECS
Sensor Cal Out Of Spec
E136 pH
The last Sensor Calibration was out of specification, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
E137
pH
Sensor User Offset At Limit
The last Sensor Offset Calibration was out of limits, check sensor condition and
connections and repeat calibration. If the message persists please consult with
your supplier.
Summary of Contents for MTD75
Page 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Page 2: ......
Page 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Page 6: ...Preface 4 MTD75 Operation Guide ...
Page 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Page 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Page 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Page 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Page 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Page 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Page 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Page 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Page 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Page 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Page 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Page 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Page 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Page 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Page 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Page 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Page 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Page 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Page 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Page 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Page 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Page 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Page 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Page 260: ...Notes 258 MTD75 Operation Guide ...
Page 261: ...Notes MTD75 Operation Guide 259 ...
Page 265: ......