pH Calibration
MTD75 Operation Guide
- 125 -
pH Cali
bration
pH Input Channel Calibration
Calibration Procedures
Normal good practices should be observed when calibrating electrode systems against standard
solutions.
Always clean the electrode system before inserting in the standard solution to avoid contamination
and to obtain the best response from the electrode system.
Rinse the electrode system in clean (preferably demineralised) water between standard solutions.
Allow time for the electrode system temperature to stabilise in each standard solution.
Use standard solutions of known quality. If in doubt make up fresh solutions.
Use clean beakers and bottles for standard solutions.
The sensor can be taken offline, which de-energises the relays and holds the current outputs at their
last values. This facility is useful when calibrating the system, however the operator must ensure the
relays are in a safe state when using this feature.
pH Buffers
Please see page 137 for information regarding the range of buffer solutions LTH can supply.
Calibration Principle
The MXD70 Series offers two different but simple methods of calibration for pH electrode systems.
Auto
In Auto calibration the instrument requires that the electrode is first placed in a 4pH buffer and then
secondly in a 9pH buffer. These are nominal values for which the instrument carries the exact LTH
buffer values due to temperature variations (other pH buffer solution values at temperature can also
be entered). From these two points the instrument then calculates the Offset and Slope for the
electrode.
In order for the auto calibration to work correctly, the buffer temperature must either be measured by
the instrument during calibration, or if manual temperature compensation is being used the buffer
temperature must be entered in the Calibration Manual Temperature Input in the calibration menu.
Manual
In Manual calibration mode it is possible to do single or two point calibration, using the
Buffer
and
Slope
menu items. It is important to do the calibration in the correct order.
1. Adjustment of the Buffer (Offset) value at 7.00pH
2. Adjustment of the Slope (Gain) value at any other pH (usually 4 or 9 pH).
The output from an ideal glass/reference electrode pair will normally be 0mV at 7pH, therefore the
slope will have little or no effect at 7pH but increasing influence the further from 7pH the calibration
point is.
If a single point calibration is required the buffer should be adjusted and no adjustment made to the
slope that will remain at the previously entered value. As actual pH buffer values are used no
compensation is made for the buffer solution with temperature in the instrument. It is important
therefore to note the actual buffer value at the temperature of the solution, see page 137 for a table
showing pH variations versus temperature for LTH’s standard solutions.
Temperature has an effect on the output from the pH electrode as well, so it is important that the
buffer temperature is entered into the Calibration Manual Temperature Input if manual temperature
compensation is being employed. Alternatively any automatic temperature compensation element
should be placed in the buffer solution with the pH sensor if Auto TC is being used.
Summary of Contents for MTD75
Page 1: ...MTD75 Cooling Tower Controller Operation Guide ...
Page 2: ......
Page 5: ...Preface MTD75 Operation Guide 3 Declaration of Conformity ...
Page 6: ...Preface 4 MTD75 Operation Guide ...
Page 12: ...Introduction 10 MTD75 Operation Guide Introduction MXD75 Termination Overview ...
Page 29: ...Installation MTD75 Operation Guide 27 Installation Horizontal Mounting Vertical Mounting ...
Page 30: ...Installation 28 MTD75 Operation Guide Installation Blank ...
Page 72: ...Biocide A 70 MTD75 Operation Guide Biocide A BLANK ...
Page 74: ...Biocide B 72 MTD75 Operation Guide Biocide B BLANK ...
Page 88: ...Inhibitor 86 MTD75 Operation Guide Inhibitor BLANK ...
Page 98: ...pH Control 96 MTD75 Operation Guide pH Control BLANK ...
Page 100: ...Alarm Relay 98 MTD75 Operation Guide Alarm Relay BLANK ...
Page 108: ...Electrodeless Conductivity Setup 106 MTD75 Operation Guide ECS Setup BLANK ...
Page 112: ...pH Input Setup 110 MTD75 Operation Guide pH Setup BLANK ...
Page 118: ...Redox Calibration 116 MTD75 Operation Guide Redox Calibration BLANK ...
Page 140: ...pH Calibration 138 MTD75 Operation Guide pH Calibration BLANK ...
Page 142: ...Calibration Reset 140 MTD75 Operation Guide Calibration Reset BLANK ...
Page 148: ...Current Outputs 146 MTD75 Operation Guide Current Outputs BLANK ...
Page 152: ...Digital Inputs 150 MTD75 Operation Guide Digital Inputs BLANK ...
Page 158: ...Configuration 156 MTD75 Operation Guide Configuration BLANK ...
Page 160: ...Update Software 158 MTD75 Operation Guide Update Software BLANK ...
Page 164: ...Save Restore Reset 162 MTD75 Operation Guide Save Restore Reset BLANK ...
Page 166: ...Error Messages 164 MTD75 Operation Guide Error Messages BLANK ...
Page 172: ...Modbus RS485 170 MTD75 Operation Guide Modbus RS485 BLANK ...
Page 184: ...Data Logging 182 MTD75 Operation Guide Data Logging BLANK ...
Page 246: ...Appendix F 244 MTD75 Operation Guide Appendix F Blank ...
Page 254: ...Appendix G 252 MTD75 Operation Guide Appendix G BLANK ...
Page 260: ...Notes 258 MTD75 Operation Guide ...
Page 261: ...Notes MTD75 Operation Guide 259 ...
Page 265: ......