71
can quit this procedure at any time.
Note
: It is very important to remove the counting slide prior to start the calibration step 1. If the
calibration step 1 is performed without removing the counting slide, cell counting result will be
inaccurate.
<Calibration Step 1>
Note
: Never turn off the Luna-FL™ cell counter during calibration, which could lead to a significant
technical failure.
7.3.4 “Calibration Step 2” is necessary for the “Custom Staining Calibration.” From this custom
calibration, users can use their own stains for cell counting. This is related to Section 2.6 Counting
Options.