21
displayed.. In the fluorescence cell counting,, bright field, green,
and red images will be displayed.
Report
The PDF report containing counting results and histograms
2.5.3 The “Auto-header” which is red-lined in the above “Save Options” image is the newly created
folder name or file name that all the images start with during the saving process. After saving, users
can see a folder or a file name, containing the saved item, which begins with “Auto-header” in the
root directory of the USB drive. In case users want to have another default name, the “Auto-header”
can be customized by editing the default value of “Logos Biosystems-”.
2.6 Counting Options
2.6.1 The “Counting Options” applies only to the bright field cell counting. Before selecting an option,
please make sure that the “Bright Field Cell Counting” is selected from the options on the Home
Screen.
2.6.2 In the “Counting Options”, users can choose one of the trypan blue staining options as below.
<Staining Options>
Staining Options
Description
With trypan blue (1:1)
This option can be used for normal bright field counting, when cell
samples are mixed with trypan blue dyes in a 1:1 ratio. This option can
generate cell viability data.