SCPC System • Safety
CP_SEM_60-03-00.002
9
Instructions on repair and replacement
!
◗
Modifications or extensions to the system or the use of spare parts which have not been tested and
approved by LIVANOVA DEUTSCHLAND GMBH may have a detrimental effect on the safety and
operation of the system. In such cases, LIVANOVA DEUTSCHLAND GMBH cannot accept any liability
or responsibility.
◗
Proof must be provided for accessories and supplementary devices which have not been tested
and approved by LIVANOVA DEUTSCHLAND GMBH that their use does not represent a safety
hazard.
◗
While the SCPC System is switched on, no electrical connections may be disconnected or
established. The SCPC System must be completely powered down when connecting or
disconnecting the SCP System.
◗
The mains switch does not disconnect the SCPC System from the mains power supply. Disconnect
the SCPC System from the power supply completely before carrying out maintenance and cleaning
work.
!
◗
Note that the SCPC System has an Uninterruptible Power Supply (UPS) which is powered by
2 batteries. When working on the device with the housing open, there is a risk of short-circuits
even if the mains cable is unplugged when the batteries are inserted and charged.
◗
Note the specific design of individual electrical components depending on the region in question,
and observe the relevant mains connection regulations according to the circuit diagram in the
Appendix section.
◗
It is essential that ESD protective safety measures are taken when working with components
susceptible to ESD. When removing or installing circuit boards or components susceptible to ESD,
ensure that you use ESD mats, wrist straps, or similar protective equipment.
◗
Observe the ESD guidelines when packing or shipping components susceptible to ESD.
◗
Always use suitable removal and insertion tools in order to avoid bending contacts (e.g. of plug
connectors) or damaging plugs or socket housings.
◗
Make sure not to damage cables or their insulation on the sharp edges of panels and circuit boards
during assembly. Route the cables in a way that prevents them from being pinched by fasteners or
by the edges of panels. Secure all cables on the cable holders provided for this purpose.
◗
A defective cable harness or a defective circuit board must not be repaired but must be replaced
completely. To guarantee safe and reliable operation of the system and the safety of patients and
users, only genuine spare parts from LIVANOVA DEUTSCHLAND GMBH must be used.
Summary of Contents for 60-03-00
Page 1: ...SCPC System Service Manual ...
Page 12: ...SCPC System Safety 12 CP_SEM_60 03 00 002 ...
Page 28: ...SCPC System System description 28 CP_SEM_60 03 00 002 ...
Page 40: ...SCPC System User menu and service menu 40 CP_SEM_60 03 00 002 ...
Page 44: ...SCPC System Error messages and error diagnosis 44 CP_SEM_60 03 00 002 ...