SCPC System • User menu and service menu
36
CP_SEM_60-03-00.002
4.2.3
Calibrate current sensor
The calibration of the current sensor does
not
form part of the regular maintenance and checking work
on the SCPC System. Re-calibration is only normally possible in the following cases:
–
After replacing the batteries
–
If the battery current display is fluctuating strongly
Calibration work should only be performed at normal room temperature (20-25 °C). To perform the
current sensor calibration:
If you close the calibration process without confirming afterwards (
Accept
), the original value saved is
retained.
◗
Press the key symbol
i-Probe Calib. Mode
➜
The calibration process is started.
Wait for approx. 2-3 minutes, before
performing the next steps.
◗
Check the display
Battery Current
(in the
Actual Values
field)
➜
If the current displayed here has
settled at a value of M ± 70 mA, it can
be assumed that the current sensor is
correctly calibrated.
◗
Press the key symbol
Accept
➜
The value determined is then
transferred to the
Stored Offset
display.
I-Probe
Calibr.
Mode
Accept