SCPC System • Error messages and error diagnosis
CP_SEM_60-03-00.002
41
5
Error messages and error diagnosis
The SCPC System is subjected to continual self-monitoring. Any errors occurring (e.g. failure of a
component) are displayed on the touch screen (both on the normal user interface and in the Service
menu).
5.1
Error messages
5.1.1
Non-specific errors
Error
Description/Possible Causes
Corrective Measures
(please observe sequence)
System will
not start
although the
batteries are
charged
Switch defective
– Check/replace switch
System will
not start if
the batteries
are not
charged or
are missing
Mains cable missing or faulty
Transformer/connecting terminal wiring
Switch defective
Mains filter defective
Rectifier faulty
– Connect/check mains cable
– Check wiring/connecting terminal
– Check/replace switch
– Check/replace mains filter
– Check/replace rectifier
System
starts
(audibly),
but display
remains dark
Contact fault (
CPU
board, display/CON 3
connection or lighting/CON 7)
– Test connector (CON 3) between
CPU
board and
display
board, clean
contacts
– Check CON 7 connector/cable, clean
contacts
– If plug contacts (CON 3) are bent or
broken, replace entire touch screen
– If the socket contacts are broken or
corroded, replace
CPU
board
Display
circuit board faulty
– Replace entire touch screen
CPU
circuit board (ZDS 2006) faulty
– Replace
CPU
board
Summary of Contents for 60-03-00
Page 1: ...SCPC System Service Manual ...
Page 12: ...SCPC System Safety 12 CP_SEM_60 03 00 002 ...
Page 28: ...SCPC System System description 28 CP_SEM_60 03 00 002 ...
Page 40: ...SCPC System User menu and service menu 40 CP_SEM_60 03 00 002 ...
Page 44: ...SCPC System Error messages and error diagnosis 44 CP_SEM_60 03 00 002 ...