SCPC System • Replacing components
56
CP_SEM_60-03-00.002
6.3.2
Removing the mains filter
It may be necessary to replace the mains switch if the SCPC System cannot be shut down. Before
removing the mains filter, the side panel (see page 6.52) needs to be removed.
fig. 27: Removing the mains filter
◗
Unfasten the two screws on the mains filter as shown
(Allen key size 3) including the washers.
◗
Unplug the five connectors from the mains filter.
Assemble the new mains filter in reverse order.