SCPC System • User menu and service menu
CP_SEM_60-03-00.002
33
4.2.1
Overview of keys and functions in the Service menu
Service menu pages 1 and 2
Function
Key symbol
Next
➜
Move between the pages in the
Service menu
Key symbol
Return
(Return key)
➜
Exit Service menu, back to start page
Service menu page 1
Key symbol
i-Probe Calib. Mode
➜
Calibration of the current sensor (e.g.
after replacing the batteries)
Key symbol
Accept
➜
Confirm the successful calibration
and apply the value determined
Key symbol
00125h > 0h
➜
Reset the hours of operation counter
Service menu page 2
Function
Key symbol
ZIS Flash Write
➜
Enable write access to the firmware
for
serial interface
(ZIS) circuit board.
The programming is carried out via an
external PC and the relevant Flash
software.
Key symbol
ZIS Reset
➜
Manual reset of the
serial interface
(ZIS) circuit board. This is also carried
out automatically when exiting the
Service menu.
I-Probe
Calibr.
Mode
Accept
00125h
=> 0h
ZIS
ZIS