Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 - Service Manual
Page 95
Rev. No 01
B. Coarse Tuning
The purpose of this procedure is used to realign the aiming position is totally out of
order. It means that the aiming is not in the center of target rod or out of the field of
view.
•
Power on the LCD touch screen first before the console
•
Double click the LIO icon and enter the pass code “777”
•
Swing the LCD panel toward to the back of the patience side
Double Clicks
'777'
Summary of Contents for TruScan 532
Page 1: ...Service Manual Multi Wavelength Pattern Scanning System LightLas 532 670 ...
Page 4: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 4 Rev No 01 ...
Page 112: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 112 Rev No 01 ...
Page 113: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 113 Rev No 01 ...
Page 114: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 114 Rev No 01 ...
Page 115: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 115 Rev No 01 ...
Page 116: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 116 Rev No 01 ...
Page 117: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 117 Rev No 01 ...
Page 127: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 127 Rev No 01 ...
Page 128: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 128 Rev No 01 ...
Page 129: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 129 Rev No 01 ...
Page 131: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 131 Rev No 01 ...
Page 140: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 140 Rev No 01 ...