Multi-Wavelength Pattern Scanning System - LightLas 532 / 670 - Service Manual
Page 12
Rev. No 01
Others
MDD 93/42/EEC (as amended by 2007/47/EC)
IEC60601-1-4 (1996) + A1 (1999)
ISO14971 (2007)
2.2
Warnings and Precautions
The following warnings and precautions apply to the Multi-Wavelength Pattern
Scanning System - LightLas 532 / 670 and should be observed by all users at all time:
DO NOT
look directly into or at the laser beam or at specular laser reflections
since direct and reflected laser light from the laser aperture can cause
permanent eye injury.
DO NOT
operate the laser unless observers are using the correct protective
eyewear. The protective eyewear must have an optical density of OD4 or
more at 532 nanometers wavelength. This information must be present on
the eyewear.
DO NOT
use objects that can readily reflect light in the vicinity of the laser
beam to avoid reflecting the beam in a hazardous manner.
DO NOT
fire the Laser directly onto flammable agents or gasses as the
focused laser beam may cause ignition.
There is no AP/ APG protection.
DO NOT
try to service or repair the laser other than what is included in this
manual. Service should only be performed by an authorized and trained
agent of the manufacturer.
DO NOT
fire the laser on a patient without first checking the operation of
the laser and verifying the optical alignment of the treatment Laser beam.
ALWAYS
use the lowest power settings possible when treating a patient
with the laser and start the treatment with minimum level of power.
ALWAYS
set the correct spot size and/or use the most appropriate one for
the power setting and type of procedure that is to be performed.
DO NOT
put the laser into ‘TREAT’ mode until ready to operate on the
patient.
DO NOT
inhale any laser plume generated by the Laser during surgery.
Personnel should take an extreme measured precaution, such as wearing
surgical masks or use plume evacuation systems when a treatment is
undergoing.
Caution - Laser plume may contain viable tissue particulates.
ALWAYS
take particular care of the optical fibers that connect the Laser
Delivery Units to the Console to make sure they do not get damaged.
Summary of Contents for TruScan 532
Page 1: ...Service Manual Multi Wavelength Pattern Scanning System LightLas 532 670 ...
Page 4: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 4 Rev No 01 ...
Page 112: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 112 Rev No 01 ...
Page 113: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 113 Rev No 01 ...
Page 114: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 114 Rev No 01 ...
Page 115: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 115 Rev No 01 ...
Page 116: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 116 Rev No 01 ...
Page 117: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 117 Rev No 01 ...
Page 127: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 127 Rev No 01 ...
Page 128: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 128 Rev No 01 ...
Page 129: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 129 Rev No 01 ...
Page 131: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 131 Rev No 01 ...
Page 140: ...Multi Wavelength Pattern Scanning System LightLas 532 670 Service Manual Page 140 Rev No 01 ...