![LI-COR LI7200RS Instruction Manual Download Page 59](http://html.mh-extra.com/html/li-cor/li7200rs/li7200rs_instruction-manual_1906669059.webp)
Note:
Some of the settings in the table below are required in order to log valid
eddy covariance datasets.
Software Field
Recommended Setting
See Section
Settings > Time
Clock Sync (PTP):
Automatic
Time Zone:
Your time zone (If your location observes
daylight savings time, choose the appropriate
offset from etc/GMT. This avoids overwriting or
creating duplicate files during a time change)
Site Setup > USB Log File
Update Rate (Hz):
10 Hz (recommended; 5 Hz min.
Note:
We
suggest setting the CO
2
/H
2
O analyzer
bandwidth at 1/2 of the logging update rate.
See
File Duration:
30 Minutes
Compress files (.ghg)
✓
(check this box)
When out of space:
As desired
Site Setup > Site Description
Station name:
Name of the station
Canopy height (m):
Measured at site
Use SmartFlux Coordinates
✓
(check this box)
Site Setup > Anemometer
Port
Port number on your SmartFlux 2 module (if
using SmartFlux 2)
Manufacturer:
Make of anemometer
Model:
Model of anemometer
Head correction applied
internally:
For the CSAT3,
✓
if true
North alignment:
For Gill models, select the anemometer setting
Wind data format:
Select the setting set in the anemometer
software
North Offset (°):
Measured at site
Height (m):
Measured at site
Table 3-2
. Checklist of mandatory and recommended EC system settings.
Section 3. Operation
3-17
Eddy Covariance Checklist
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......