Section 4. User Calibration
4-2
When the span is set, the value of
S
c0
(or
S
w0
for water) is determined. For
example, if there is a known CO
2
density
ρ
'c in the optical path and the measured
absorptance is
α
’c
, then from equations 8-19 and 8-8, we can write
so
4-3
We rewrite this in terms of a known mole fraction
C’
f
instead of density.
4-4
If we assume that the CO
2
span gas is dry, then
P
ec
=
P
, so
4-5
Similarly, for a known H
2
O mole fraction
W’
f
and resulting measured absorptance,
ɑ
'
w
,
4-6
Whenever the CO
2
span is set, the instrument saves two values that are used if and
when a secondary CO
2
span is performed. These values are
4-7
When a secondary span for CO
2
is performed at a mole fraction
C’
f
with measured
absorptance
ɑ
’
c
, then the span slope term
S
c1
is computed from
4-10
LI-7200RS Enclosed CO₂/H₂O Analyzer
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......