Section 2. Initial Setup
1
Remove the intake cap.
Carefully observe the flow arrow indicator on the filter. Air should flow in the dir-
ection of the arrow.
2
Remove the compression fitting from the intake side of the filter.
3
Install the nut, front ferrule, and back ferrule over the short stainless steel pipe.
Be sure the tube is fully seated in the filter. Tighten the fitting to hand tight, then
about ½ turn more using 9/16" and 1" wrenches.
4
Install the nut, front ferrule, and back ferrule on the intake tube.
5
Install the inlet cap.
Installing the Intake Cap
The intake cap kit includes the following components:
l
Intake Cap (includes screen and hose clamp).
l
Spare screen, retaining ring, and hose clamp.
The intake cap attaches to the standard ¼" diameter intake tube (part number 9972-
053) and heated intake tube. To install the intake cap:
2-10
LI-7200RS Enclosed CO₂/H₂O Analyzer
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......