Section 7. Software Reference
Test Point
Description
Normal range
IR A/D (V)
Voltage from IRGA being sent to
A/D converter
0.2 – 4.8 V
Detector Temp (V)
Detector temperature voltage
2.9 V. Readings near +5 V
or 0 V are normal for a
few seconds after power
on.
Aux Ref +
16-bit conversion of positive
reference voltage used in Auxiliary
Input circuit
~64830
Aux Ref -
16-bit conversion of negative
reference voltage used in Auxiliary
Input circuit
~600
ADC CH1 (V)
Analog input channel 1
-5 V to +5 V. A reading
near 0 is normal for an
open input.
ADC CH2 (V)
Analog input channel 2
-5 V to +5 V. A reading
near 0 is normal for an
open input.
ADC CH3 (V)
Analog input channel 3
-5 V to +5 V. A reading
near 0 is normal for an
open input.
7-30
LI-7200RS Enclosed CO₂/H₂O Analyzer
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......