![LI-COR LI7200RS Instruction Manual Download Page 63](http://html.mh-extra.com/html/li-cor/li7200rs/li7200rs_instruction-manual_1906669063.webp)
Note:
We suggest setting the CO
2
/H
2
O analyzer bandwidth at 1/2 of the logging
update rate. Thus, if you are sampling at 10 Hz, the bandwidth should be set to 5
Hz. 5 Hz is the default setting for bandwidth.
Entering the Site Description
The
Site Description
tab is used to enter information about the site. This inform-
ation configures the metadata file that is used for flux processing.
Site name:
Name of the research site (e.g., Marsh).
Station name:
Name of the flux stations within the site.
Canopy height
(mandatory)
:
Site canopy height; meters.
Displacement height:
Also referred to as zero plane displacement height, the dis-
placement height of a vegetated surface (usually designated
d
) is the height at
which the wind speed would go to zero if the logarithmic wind profile was main-
tained from the outer flow all the way down to the surface (that is, in the absence
of vegetation). In other words, it is the distance over the ground at which a non-
vegetated surface should be placed to provide a logarithmic wind field equal to the
observed one. For forest canopies, the displacement height is estimated to vary
between 0.6 and 0.8 times the height of the canopy. If not entered explicitly,
EddyPro computes displacement height as:
3-1
d
= 0.67 x canopy height
Section 3. Operation
3-21
Entering the Site Description
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......