Type / Element
Value / Description
Digital outputs
4
8 channels
DO 1
Valve closed
DO 2
Valve open
DO 3
Valve in position
DO 4
Valve error
DO 5
Sensor error
DO 6
ready
DO 7
Emission ON
DO 8
Sensor status
Digital output DO 3, Deviation
5
Signal ON
0.1% … 50.0% F.S. (of pressure sensor)
Signal OFF
0.1% … 50.0% F.S. (of pressure sensor)
ON-delay [s]
0 … 999
Voltage [VDC]
+24
Current [mA]
1 output max. 100
The current of all outputs together must not
exceed 150
Analog outputs
6
4 channels
AO 1
Pressure sensor signal
AO 2
7
Valve signal (0 … 10 VDC)
AO 3
Valve position MOVE 1250
AO 4
+10 VDC reference voltage (10 mA)
Voltage range [VDC]
0 … +10
Load resistance [kΩ]
≥2
4
For information on wiring see chapter «5.2.4 Automatic Booting Function», on configuration see chapter
«5.2.11 Digital Outputs», on operation see chapter «5.4 Operation Via Inputs and Outputs».
5
For explanation and information on configuration see chapter «5.2.11 Digital Outputs».
6
For information on wiring see chapter «4.2.6 Analog Inputs/Outputs», on configuration see chapter
«5.2.12 Analog Inputs/Output», on operation see chapter «5.4 Operation Via Inputs and Outputs».
7
The 0 … 10 VDC analog output AO 2 can be used for control valves other than the MOVE 1250 and
MOVE X.